FDAClass IIAugust 5, 2020

Orthopedics 6.0MM ROD recall

Orthopedics 6.0MM ROD, COCR, SINGLE HEX, 500MM LENGTH-The RESPONSE" 5.5/6.0 Spine System is intended for immobilization and stabilization of the…

Why was Orthopedics 6.0MM ROD recalled?

Product may have incorrect laser etched alignment lines, the non-conforming laser etch may have compromised strength, lead to unintended fracture of the device

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Lot Number: 108332-A UDI: (01)0084113212400(10)108332A

Product
Orthopedics 6.0MM ROD, COCR, SINGLE HEX, 500MM LENGTH-The RESPONSE" 5.5/6.0 Spine System is intended for immobilization and stabilization of the…
Recalling firm
OrthoPediatrics Corp
Quantity
73 units
Distribution
US Nationwide distribution including in the states of Al, CA, FL, GA, KY, MI, MO, NC, NY, OH, TX.
Recall date
August 5, 2020
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls