CPAP Mask Cushion recall
CPAP Mask Cushion, Model Nos. PB781S, PB781M, PB781L, 781S, 781M and 781L Product Usage: A CPAP mask is attached to an end user s face and a CPAP…
Why was CPAP Mask Cushion recalled?
A May 2015 design change in the cushion seal replacement part and accompanying elbow replacement part resulted in incompatibility of the new cushion seal with the previous design of the elbow. Use of the new cushion seal with the previous design elbow could lead to unacceptable CO2 porting. As there was no model number change for this design change, customers may not be aware of the incompatibility.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
All units manufactured after May 1, 2015
- Product
- CPAP Mask Cushion, Model Nos. PB781S, PB781M, PB781L, 781S, 781M and 781L Product Usage: A CPAP mask is attached to an end user s face and a CPAP…
- Recalling firm
- Compass Health Brands (Corporate Office)
- Quantity
- 742
- Distribution
- US nationwide distribution to AL, CT, FL, GA, KY, MS, NC, NY, TN, and TX.
- Recall date
- August 29, 2018
- Classification
- Class I
- Status
- Terminated
- Agency
- FDA (device)