FDAClass IIAugust 5, 2020

Progressa Bed recall

Progressa Bed, Catalog No. P7500 - Product Usage: intended to be used to treat or prevent pulmonary or other complications associated with immobility…

Why was Progressa Bed recalled?

Affected beds may have screws installed that do not meet material tensile strength requirements. Failure of this screw could result in serious patient or user injury.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Serial Numbers Affected: V133AW9885 through V142AW0261and manufactured between May 12, 2020 and May 21, 2020

Product
Progressa Bed, Catalog No. P7500 - Product Usage: intended to be used to treat or prevent pulmonary or other complications associated with immobility…
Recalling firm
Hill-Rom, Inc.
Quantity
257
Distribution
Worldwide distribution - US nationwide distribution including in the states of AR, AZ, CA, DE, IN, KY, MD, ME, MI, MO, NC, NM, NY, OH, OK, PA, SC, TN, TX, VA, WA, WI, WV and the country of Canada.
Recall date
August 5, 2020
Classification
Class II
Status
Terminated