BrightView X recall: Software Issue
BrightView X, model #882478; gamma camera designed for single or dual detector nuclear imaging
Why was BrightView X recalled?
The firm became aware of an issue where during a non-Auto Body Contouring (ABC), non-circular, rel-180 scan, the detector may come in contact with the patient. The issue was caused by a software defect in the workflow setup script.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
11000003, 11000008, 11000030, 11000004, 11000005, 11000006, 11000007, 11000009, 11000010, 11000011, 11000012, 11000013, 11000014, 11000016, 11000017, 11000018, 11000019, 11000020, 11000022, 11000023, 11000024, 11000025, 11000026, 11000027, 11000028, 11000029, 11000031, 11000032, 11000033, 11000034, 11000035, 11000036, 11000037, 11000038, 11000039, 11000040, 11000041, 11000042, 11000043, 11000044, 11000045, 11000046, 11000047, 11000048, 11000049, 11000050, 11000051, 11000052, 11000054, 11000055, 11000056, 11000057, 11000058, 11000059, 11000060, 11000061, 11000062, 11000063, 11000064, 11000065, 11000066, 11000067, 11000068, 11000069, 11000070, 11000071, 11000072, 11000073, 11000074, 11000075, 11000076, 11000077, 11000078, 11000079, 11000080, 11000081, 11000083, 11000084, 11000085, 11000086, 11000087, 11000088, 11000089, 11000090, 11000091, 11000092, 11000093, 11000094, 11000095, 11000096, 11000097, 11000098, 11000099, 11000100, 11000101, 11000102, 11000103, 11000104, 11000105, 11000106, 11000107, 11000108, 11000109, 11000110, 11000112, 11000113, 11000114, 11000115, 11000116, 11000117, 11000118, 11340001, 11340002, 11340003, 11340004, 11340005, 11000015
- Product
- BrightView X, model #882478; gamma camera designed for single or dual detector nuclear imaging
- Recalling firm
- Philips Medical Systems (Cleveland) Inc
- Quantity
- 117 units
- Distribution
- Worldwide distribution. US nationwide, Algeria, Argentina, Armenia, Australia, Austria, Belgium, Brazil, Czech Republic, Denmark, Finland France, Gabon, Germany, Greece, India, Indonesia, Ireland, Japan, Korea, Republic of Korea, Republic of Lebanon, Lithuania, Malaysia, Netherlands, New Zealand, Norway, Oman, Panama, Philippines, Poland, Portugal, Russian Federation, Saudi Arabia, Slovakia, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, Viet Nam.
- Recall date
- September 14, 2016
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)