FDAClass IIOctober 8, 2014

Verigene BC-GP Kit recall

Verigene BC-GP Kit (Catalog number 20-005-018); 1) 20 Verigene BC-GP Test Cartridges; are qualitative multiplexed in vitro diagnostic tests for…

Why was Verigene BC-GP Kit recalled?

There is a specific Extraction Tray lot containing Tips that may slightly increase the occurrence of Process SP Processing Error when used due to the Processor SP experiencing difficulty releasing the Tip to the final location in the Tip Assembly. The difficult releasing the Tip can result in a Processing Error and occasionally the Tip to be dropped into the Processor SP.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Part Number: 20-009-018; Lot Numbers: 051713018B, 052413018C

Product
Verigene BC-GP Kit (Catalog number 20-005-018); 1) 20 Verigene BC-GP Test Cartridges; are qualitative multiplexed in vitro diagnostic tests for…
Recalling firm
Nanosphere, Inc.
Quantity
1,660 extraction trays
Distribution
Worldwide Distribution:US (Nationwide) including states of: AZ, CA, FL, IL, KY, MD, MN, MT, OH, PA, TN, TX, VA, WA, WI; and countries of: Belgium, China, and Saudi Arabia.
Recall date
October 8, 2014
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls