FDAClass IAugust 12, 2020

Alaris PCA Module recall

Alaris PCA Module Model 8120, a modular infusion pump and monitoring system

Why was Alaris PCA Module recalled?

Cleaning agents may damage Inter-Unit Interface (IUI) Connectors, which may lead to interruption of communication or power between PC Unit and modules, which may result in an infusion that stops with a CHANNEL DISCONNECTED message, and an audio/visual alarm on the PC Unit and an interruption of therapy or monitoring.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

During cleaning, updated cleaning instructions and IUI covers must be used for all serial numbers.

Product
Alaris PCA Module Model 8120, a modular infusion pump and monitoring system
Recalling firm
CareFusion 303, Inc.
Quantity
83532
Distribution
Worldwide distribution. US nationwide including Guam, Puerto Rico, American Samoa, and Military Pacific. Countries of: TR, QA, CA, AE, KW, MX, GI, IL, SA, CH, NL, BE, TW, ZA, PH, CN, AU, IN, FR, MY, NZ, GB, KR, PK, DE, SG, JP, CU, IT, and BH
Recall date
August 12, 2020
Classification
Class I
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls