FDAClass IIOctober 8, 2014

Siemens Dimension Vista¿ CHEM 1 CAL recall

Siemens Dimension Vista¿ CHEM 1 CAL, lot 3GM081. The CHEM 1 CAL is an in vitro diagnostic product for the calibration of Calcium (CA), Cholesterol…

Why was Siemens Dimension Vista¿ CHEM 1 CAL recalled?

Siemens Healthcare Diagnostics has received complaints for low recovery of Magnesium (MG) Quality Control (QC) and patient samples following calibration of MG with the Dimension Vista® CHEM 1 CAL, lot 3GM081. Investigation by Siemens confirms that QC and patient samples under-recovers MG by 0.25 mg/dL [0.10 mmol/L] across the assay range following calibration with 3GM081.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

lot 3GM081, exp. 07-01-2014

Product
Siemens Dimension Vista¿ CHEM 1 CAL, lot 3GM081. The CHEM 1 CAL is an in vitro diagnostic product for the calibration of Calcium (CA), Cholesterol…
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Quantity
9769 Total (6000 US)
Distribution
Worldwide Distribution - USA (Nationwide) and Internationally to Australia Austria Belgium Canada Czech Republic Finland France Germany Great Britain Italy Japan Netherlands New Zealand Norway Portugal Slovak Republic Slovenia South Korea Spain Switzerland
Recall date
October 8, 2014
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls