FDAClass IISeptember 14, 2016

Merge RadSuite software. Radiological image processing recall

Merge RadSuite software. Radiological image processing system.

Why was Merge RadSuite software. Radiological image processing recalled?

When RadSuite is used with IPID (Issuer of Patient ID) as a part of the "Patient Identifier," it is possible in some circumstances that the demographics of one patient will be applied to a study or studies for another patient.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

RadSuite Versions: 5.30.8, 5.35.3, 5.35.4, 5.35.5, 8.30.7.3, 8.30.7.4, 8.30.7.5, 8.30.7.7 are affected, BUT ONLY if they are configured with all of the following conditions: (1) Must have IPID enabled; (2) Must use IPID morphers (3) Must Define a Defaultvalue in the IPID Morpher; (4) Must use the Aggregating Morpher; (5) Must not set OverwritelfPresent=false; and (6) Must be on an internal store (this is data that is already present on EA being stored to Radsuite, this excludes direct DICOM stores to the EA).

Product
Merge RadSuite software. Radiological image processing system.
Recalling firm
Merge Healthcare, Inc.
Quantity
10 sites have the affected software that is configured with the specific conditions listed in Code Information
Distribution
Distribution was made to medical facilities in AL, MI, MO, PA, TN, and TX.
Recall date
September 14, 2016
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls