Spacelabs Medical Patient Monitor is a Physiological recall
Spacelabs Medical Patient Monitor is a Physiological, Patient monitor With Arrhythmia Detection or Alarms. The Models are 91367, 91369, 91370, 91387…
Why was Spacelabs Medical Patient Monitor is a Physiological recalled?
The clinician did not realize the monitor was in END CASE mode and was expecting audible alarms. Visual alarms continue to work normally. The firm performed a retrospective analysis of all recalls since 2011 and discovered 222 additional serial numbers were missed and were not included in the previous recall under RES 59456 (recall number Z-3032-2011).
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
The same KIT PART Number (PN 040-1458-00) identified on 12 units distributed IN the US and on 3 units distributed OUTSIDE of the US. These 15 units will receive serial numbers after the software upgrade is installed. THERE ARE 70 SERIAL NUMBERS ON UNITS DISTRIBUTED IN THE US: 040-1548-00, 1367-000033, 1369-001517, 1369-003051, 1369-003077, 1369-004247, 1369-005914, 1369-006877, 1369-007165, 1369-008519, 1369-010292, 1369-011637, 1369-017520, 1369-061217, 1370-000013, 1370-000014, 1370-000025, 1370-000095, 1370-000285, 1370-000449, 1370-000471, 1370-000579, 1370-000658, 1370-001700, 1370-002494, 1370-002510, 1370-003785, 1370-004176, 1370-004184, 1370-004275, 1370-004292, 1387-000465, 1387-000535, 1387-000543, 1387-001071, 1387-001677, 1387-006200, 1387-010824, 1387-010830, 1387-011914, 1387-014142, 1387-014159, 1387-014184, 1387-014186, 1387-014187, 1387-014188, 1387-014596, 1387-015721, 1387-015722, 1387-015723, 1387-016045, 1387-018424, 1387-018947, 1387-019766, 1387-019768, 1387-019769, 1387-019771, 1387-019772, 1387-100036, 1387-101646, 1387-104672, 1387-105801, 1387-105811, 1387-105812, 1387-105882, 1393-000442, 1393-000579, 1393-000580, 1393-000581, 1393-000582, and 387-002421. THERE ARE 137 SERIAL NUMBERS IDENTIFIED ON UNITS DISTRIBUTED OUTSIDE OF THE US: 1369-000011, 1369-000165, 1369-000407, 1369-000928, 1369-002976, 1369-003388, 1369-005316, 1369-005334, 1369-006100, 1369-006281, 1369-007338, 1369-009142, 1369-009143, 1369-009144, 1369-009145, 1369-009146, 1369-009889, 1369-009890, 1369-011741, 1369-011742, 1369-011743, 1369-011744, 1369-011745, 1369-013721, 1369-016648, 1369-016649, 1369-017759, 1369-017791, 1369-017792, 1370-000026, 1370-000725, 1370-000771, 1370-002373, 1370-002378, 1370-002380, 1370-002381, 1370-002382, 1370-002383, 1370-002384, 1370-003809, 1370-004133, 1370-004277, 1370-004509, 1370-004510, 1370-004511, 1370-004516, 1370-004617, 1370-200366, 1370-200370, 1370-200374, 1370-200375, 1370-200376, 1370-200378, 1370-200380, 1370-200471, 1370-200473, 1370-200474, 1370-200475, 1370-200476, 1370-200691, 1387-000044, 1387-000447, 1387-000467, 1387-000548, 1387-001086, 1387-002157, 1387-005006, 1387-008003, 1387-008310, 1387-010742, 1387-013241, 1387-013716, 1387-014610, 1387-018104, 1387-018105, 1387-018171, 1387-018176, 1387-018277, 1387-018638, 1387-018817, 1387-019563, 1387-100148, 1387-100150, 1387-100151, 1387-100155, 1387-100906, 1387-101061, 1387-101062, 1387-101063, 1387-101064, 1387-101065, 1387-101066, 1387-101067, 1387-102222, 1387-103550, 1387-103659, 1387-104462, 1387-105831, 1387-105832, 1387-105833, 1387-105834, 1393-000013, 1393-000517, 1393-000518, 1393-000519, 1393-000520, 1393-000521, 1393-000522, 1393-000523, 1393-000524, 1393-000525, 1393-000808, 1393-000809, 1393-000822, 1393-000831, 1393-000833, 1393-000834, 1393-000851, 1393-000852, 1393-00235, 1393-00236, 1393-00237, 1393-00238, 1393-00239, 1393-00240, 1393-00241, 1393-00335, 1393-00346, 1393-00347, 1393-00348, 1393-00349, 1393-00350, 1393-00351, 1393-00352, 1393-00353, 1393-00354, AND 1393-00355.
- Product
- Spacelabs Medical Patient Monitor is a Physiological, Patient monitor With Arrhythmia Detection or Alarms. The Models are 91367, 91369, 91370, 91387…
- Recalling firm
- Spacelabs Healthcare Inc
- Quantity
- 82 units distributed in US and 140 units distributed outside US
- Distribution
- Worldwide Distribution- US (nationwide) including the states of: FL, GA, IA, KY, LA, MA, MI, MN, MO, MS, NJ, NY, OH, OK, PA, SC, TN, TX, WA, Guam and Puerto Rico and to the following countries: Argentina, Australia, Canada, China, Czech Republic, Finland, France, Germany, Greece, India, Israel, Italy, Netherlands, New Zealand, Panama, Paraguay, Poland, Singapore, Spain, Sri Lanka, Switzerland, Tunisia, United Kingdom, and Uruguay.
- Recall date
- October 1, 2014
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)