FDAClass IIJuly 19, 2017

Leksell Gamma Knife Icon is a teletherapy device recall

Leksell Gamma Knife Icon is a teletherapy device intended for stereotactic irradiation of head structures ranging from very small target sizes of a…

Why was Leksell Gamma Knife Icon is a teletherapy device recalled?

If the gantry module is replaced after original installation the configuration settings may be missing. The identified risk for this issue is electrical safety for technicians doing maintenance on the X-ray generator.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Serial Numbers 6018, 6047, 6090, 6094, 6171, 8018

Product
Leksell Gamma Knife Icon is a teletherapy device intended for stereotactic irradiation of head structures ranging from very small target sizes of a…
Recalling firm
Elekta, Inc.
Quantity
2
Distribution
US Distribution to the states of : NY and PA.
Recall date
July 19, 2017
Classification
Class II
Status
Terminated
Agency
FDA (device)