FDAClass IJuly 22, 2026

ABIOMED Impella CP Set recall

ABIOMED Impella CP Set, Pump Set Code 0048-0044; containing Introducer Product Code 0052-3046. heart pump and accessories

Why was ABIOMED Impella CP Set recalled?

A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers are provided in every pump set.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Pump Set Code 0048-0044; Introducer Product Code 0052-3046. GTIN: 00813502012200. All current batches in field.

Product
ABIOMED Impella CP Set, Pump Set Code 0048-0044; containing Introducer Product Code 0052-3046. heart pump and accessories
Recalling firm
Abiomed, Inc.
Quantity
557 units
Distribution
Worldwide distribution. US Nationwide, Australia (AU), Austria (AT), Belgium (BE), Brazil (BR), Brunei (BN), Canada (CA), China (CN), Croatia (HR), Czech Republic (CZ), Denmark (DK), Finland (FI), France (FR), Germany (DE), Greece (GR), Hong Kong (HK), Ireland (IE), Israel (IL), Italy (IT), Japan (JP), Kuwait (KW), Luxembourg (LU), Malaysia (MY), Mexico (MX), Netherlands (NL), Norway (NO), Panama (PA), Poland (PL), Saudi Arabia (SA), Serbia (RS), Slovenia (SI), South Korea (KR), Spain (ES), Sweden (SE), Switzerland (CH), Taiwan (TW), Thailand (TH), United Arab Emirates (AE), and United Kingdom (GB).
Recall date
July 22, 2026
Classification
Class I
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls