FDAClass IIAugust 31, 2016

TrueBeam¿ Radiotherapy Delivery System and EDGE" recall

TrueBeam¿ Radiotherapy Delivery System and EDGE" Radiotherapy Delivery System, K140528. The TrueBeam and Edge Systems are intended to provide…

Why was TrueBeam¿ Radiotherapy Delivery System and EDGE" recalled?

Vision RT Ltd received three (3) reports from different sites that discovered the following issue, unintended changes were made to planned couch parameters, specifically the couch rotation parameter, during patient set-up for the Optical Surface Monitoring System [OSMS]. 1. CRM 3194, No Serious Injury, No MDR, Aware date: Dec 11, 2015 2. CRM 3321, No Patient Involved, No MDR, Aware date: Jan 19, 2016 3. CRM 4570, No Serious Injury, No MDR, Aware date: March 15, 2016 No patient harm was reported in any of these cases.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Serial Number of the device(s) distributed 248-016 248-021 248-022 248-030 248-038 248-041 248-063 248-064 249-1514 249-1516 248-071 249-0239 248-073 248-080 248-084 249-0043 249-0055 249-0056 249-0058 249-0069 249-0130 249-0137 249-1509 249-0141 249-0275 249-0152 249-0162 249-0169 249-0251 249-0253 249-0254 249-0288 249-1503 249-1512 249-1518 249-TBD 249-0061 249-0131 249-0256 249-0289 249-0290 249-TBD 248-083 249-0273 249-0276 249-1500 249-1519

Product
TrueBeam¿ Radiotherapy Delivery System and EDGE" Radiotherapy Delivery System, K140528. The TrueBeam and Edge Systems are intended to provide…
Recalling firm
Vision RT Ltd
Quantity
47
Distribution
Product distributed in US. The foreign countries where the product was distributed are as follows: Canada, Germany, Japan, Sweden, Switzerland, Taiwan, and United Kingdom. No product was distributed to US government accounts.
Recall date
August 31, 2016
Classification
Class II
Status
Terminated
Agency
FDA (device)