FDAClass IIOctober 11, 2023

ADVIA Chemistry Iron_2 recall

ADVIA Chemistry Iron_2 (IRON_2) Reagents, Material Numbers 10377510 (6 x 350 tests) and 10341118 (7 x 145 tests)

Why was ADVIA Chemistry Iron_2 recalled?

There is a potential for a positive bias on Quality Control (QC) and patient sample results when using some IRON_2 reagent wedge pairs. When the issue occurs, the observed bias will vary wedge to wedge.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

(6 x 350 tests) UDI-DI: 00630414561974; (7 x 145 tests) UDI-DI: 00630414513157; Lot Numbers: 612584 625096 612579 625091 635826 635821

Product
ADVIA Chemistry Iron_2 (IRON_2) Reagents, Material Numbers 10377510 (6 x 350 tests) and 10341118 (7 x 145 tests)
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Quantity
4966 reagents
Distribution
US Nationwide. Global Distribution
Recall date
October 11, 2023
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls