Veran : Endobronchial recall
Veran : Endobronchial Models: INS-5500 View Optical Probe INS-5450 SPiN Xtend¿ 21ga, 1.8 OD, 5/Bx INS-5440 View Peripheral Catheter, 3.8mm OD, 2.5mm…
Why was Veran : Endobronchial recalled?
Non-conformances in Veran devices (ig4 Image Guided System, SPiN Thoracic Navigation System, SPiN Planning Laptop Workstation, and all associated navigation instruments) manufacturing to design requirements and insufficient evidence to demonstrate adequate electromagnetic compatibility (EMC). EMC is the ability of electrical equipment and systems to function acceptably without interference and to not interfere with other devices within its operating environment
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Model-UDI: INS-5500 View Optical Probe 00815686020729; INS-5450 SPiN Xtend¿ 21ga, 1.8 OD, 5/Bx 00815686020651; INS-5440 View Peripheral Catheter, 3.8mm OD, 2.5mm WC 00815686020644; INS-5411 0.016" Stylet, 1235 mm L, 10/Bx 00815686021191; INS-5410 SPiN Flex¿ ANSO¿ Needle, 22ga, 1.8mm OD, 5/Bx 00815686020637; INS-5323 Sliding stopper 00815686020712; INS-5300 "Always-On Tip Tracked¿ Triple Needle Brush, 12mm L, 1.8mm OD, 5/Bx" 00815686021207 . All serial numbers and all lot numbers
- Product
- Veran : Endobronchial Models: INS-5500 View Optical Probe INS-5450 SPiN Xtend¿ 21ga, 1.8 OD, 5/Bx INS-5440 View Peripheral Catheter, 3.8mm OD, 2.5mm…
- Recalling firm
- Olympus Corporation of the Americas
- Quantity
- 6318 units
- Distribution
- Worldwide distribution - US Nationwide and the countries of CN, DE, HK, IT, KR, SG, TH, TW.
- Recall date
- October 11, 2023
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)