Medfusion Syringe Infusion Pump recall
Medfusion Syringe Infusion Pump, containing the Barrel Clamp Guide mechanism
Why was Medfusion Syringe Infusion Pump recalled?
Certain Barrel Clamp Guides may contain a ridge that could potentially lead to spring slippage, resulting in the inability of the pump to recognize a loaded syringe or the pump, misidentifying the size of a loaded syringe.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Barrel Clamp Guide mechanism number: G6000716; Barrel Clamp Guide lot numbers: P0239588, P0321444, and P0462421
- Product
- Medfusion Syringe Infusion Pump, containing the Barrel Clamp Guide mechanism
- Recalling firm
- Smiths Medical ASD Inc.
- Quantity
- 276 units
- Distribution
- US distribution to AL, AZ, CA, DC, FL, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, NJ, NM, NY, OH, OK, PA, TN, TX, UT, VA, WA. International distribution to Canada and Lebanon.
- Recall date
- August 15, 2018
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)