FDAClass IIJuly 12, 2017

Logic Fit Tibial Tamp Head recall: Labeling Error

Logic Fit Tibial Tamp Head

Why was Logic Fit Tibial Tamp Head recalled?

The firm determined that the labeling (surgical technique) should be updated to clarify the proper technique and instrumentation to remove the tamp assembly from the bone, i.e., to include a caution statement about the potential for instrument breakage if the tamp handle/guide is misused by impacting in retro-grade.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Catalog #: 213-75-01 Lot #'s: 37166-002, 44944-004, 44944-005, 44944-006, 48073-010, 48073-011, 48073-012, 49036-017, 51180-004, 51180-005, 51180-006, 51180-008, 51754-031, 51754-032, 51754-033, 51754-034, 59743-001, 61872-001,and 61872-002.

Product
Logic Fit Tibial Tamp Head
Recalling firm
Exactech, Inc.
Quantity
573
Distribution
AL, AZ, CA, CA, CO, FL, GA, IA, ID, IL, IN, KS, LA, MA, MD, ME, MN, NC, NJ, NV, NY, OH, OK, OR, RI, SC, TN, TX, VA, WA, WI, Hawaii and Puerto Rico Australia, France, Germany, India, Italy, Japan, Korea, Luxembourg, Singapore, Spain, Switzerland, The Netherlands and United Kingdom
Recall date
July 12, 2017
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls