FDAClass IIAugust 31, 2016

Ingenuity Core recall: Software Issue

Ingenuity Core, Model number 728321 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional images of the body by…

Why was Ingenuity Core recalled?

After upgrading to 3.6.7 software version via FCO72800643, during reconstruction of gated helical scans, the planned anatomy can be cut off on the CT images. This issue can occur on gated CT helical reconstruction that is not planned at iso-center (0,0). If the region of interest is not visualized in the images, a CT rescan may be performed.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

728321 -- Ingenuity Core 310004 310005 310012 310016 310020 310021 310027 310028 310033 310034 310041 310044 310045 310057 310061 310065 310072 310074 310076 310077 310078 310094 310099 310100 310103 310105 310117 310120 310122 310135

Product
Ingenuity Core, Model number 728321 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional images of the body by…
Recalling firm
Philips Medical Systems (Cleveland) Inc
Quantity
30
Distribution
US Nationwide Distribution in the states of: AL,AR,AZ,CA,CO,CT,DC,DE,FL,GA,HI,IA,ID,IL,IN,KS,KY,LA,MA, MD,ME,MI,MN,MO,MS,NC,ND,NE,NJ,NM, NV,NY,OH,OK,OR,PA,PR,SC,TN,TX,UT,VA,VT,WA,WI,WV,WY
Recall date
August 31, 2016
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls