FDAClass IISeptember 24, 2014

TRx Vertical Positioner recall

TRx Vertical Positioner, Classification Name: Wheelchair Component - Power Positioning System.

Why was TRx Vertical Positioner recalled?

Motion Concepts identified that there is a possibility of knee support failure occurring through fatigue of two attachment bolts in the Vertical Position System (TRx Vertical Positioner).

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

All units manufactured between June 2011 and May 2014: Serial Numbers 28583, 28597, 28604, 28607, 28629, 30334, 028380, U001050, U001083, U001544, U002020, U002326, U002386, U002403, U002411, U002489, U002502, U002505, U002509, U002510, U002519, U002716, U002922, U003577, U004011, U004018, U004073, U004079, U004145, U004214, U004400, U004404, U004472, U004485, U006042, U006234, U006235, U006260, U006295, U006296, U006353, U006378, U007126, U007138, U007163, U007239, U007245, U007279, U007288, U007407, U008556, U008631, U008698, U008840, U008896, U009548, U009569, U009579, U009707, U009718.

Product
TRx Vertical Positioner, Classification Name: Wheelchair Component - Power Positioning System.
Recalling firm
Motion Concepts
Quantity
US: 59 units
Distribution
US Distribution in states of: AR, CA, CT, DE, IA, ID, IL, IN, MI, MD, MO, NE, NY, OH, PA, TN, TX, and WA.
Recall date
September 24, 2014
Classification
Class II
Status
Terminated
Agency
FDA (device)