FDAClass IISeptember 24, 2014

VERO Linear Accelerator System recall: Software Issue

VERO Linear Accelerator System, Model Number MHI-TM2000; Common Name: Medical Linear Accelerator. MHI-TM2000 Linear Accelerator System is intended for…

Why was VERO Linear Accelerator System recalled?

Software Anomaly: If a user changes the calendar setting from Workday to Holiday or vice versa, that would erroneously change the status of treatment/fractions, which are completed and to become Treated (completed) status on the exact day when such change is made, to Untreated status. Furthermore, such treatment would be cloned and mistakenly added to the schedule as Untreated treatment

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Software Version 3.0.0 and after. Serial Numbers 201902, 203901, and 203919

Product
VERO Linear Accelerator System, Model Number MHI-TM2000; Common Name: Medical Linear Accelerator. MHI-TM2000 Linear Accelerator System is intended for…
Recalling firm
MITSUBISHI HEAVY INDUSTRIES, LTD., HIROSHIMA MACHINERY WORK
Quantity
3 units in use in the US
Distribution
Distributed in the states of NY, FL, and TX.
Recall date
September 24, 2014
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls