AXIOM Luminos dRF system with SW VD10 or on Luminos dRF Max The Axiom Luminos dRF is intended to be used as a universal diagnostic imaging system for…
Why was AXIOM Luminos dRF system with SW VD10 or on Luminos dRF recalled?
It was determined that a potential malfunction may occur on the Luminos dRF system with SW VD10 or on Luminos dRF Max when the table side control is exposed to fluids. Fluid may infiltrate the tableside control panel and result in malfunction. In the very unlikely case of fluid infiltration a permanent short circuit within the control panel of the Luminos dRF / Luminos dRF Max will automatically d
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
model # 10094200, see distribution list for individual device serial numbers
- Product
- AXIOM Luminos dRF system with SW VD10 or on Luminos dRF Max The Axiom Luminos dRF is intended to be used as a universal diagnostic imaging system for…
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Quantity
- 33
- Distribution
- Nationwide Distribution including IL VA TN PA TX IN MN NM GA MI FL CA NY ME OK MO NJ and AZ.
- Recall date
- September 24, 2014
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)