Cardiosave Hybrid Intra-Aortic Balloon Pump recall
Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), Model Nos. 0998-00-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-00-0800-34, 0998-00-0800-35…
Why was Cardiosave Hybrid Intra-Aortic Balloon Pump recalled?
The Cardiosave IABP may unexpectedly shut down when the device is running on AC power, only one battery is installed in the IABP, and the battery is physically removed while the battery is being charged. This occurs during a very specific set of conditions.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
All lots
- Product
- Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), Model Nos. 0998-00-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-00-0800-34, 0998-00-0800-35…
- Recalling firm
- Datascope Corp.
- Quantity
- 3969 total (US), 3550 total (OUS)
- Distribution
- Nationwide domestic distribution. Foreign distribution worldwide.
- Recall date
- October 6, 2021
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)