FDAClass IISeptember 27, 2023

Prucka 3 Amplifiers recall

Prucka 3 Amplifiers, Model P1801PA, and Field Replaceable Units (FRUs), Model 5875569, used with CardioLab AltiX and ComboLab AltiX.

Why was Prucka 3 Amplifiers recalled?

A diode on the power supply of the Prucka 3 Amplifier used with CardioLab/ComboLab systems could reach elevated temperatures and fail, then power off and become inoperable.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

GTIN 00195278507044 (CardioLab) and 00195278507051 (ComboLab) Serial numbers for Amplifiers: SVS22330007SA, SVS22330008SA, SVS22330009SA, SVS22330010SA, SVS22330011SA, SVS22330012SA, SVS22330013SA, SVS22330014SA, SVS22330016SA, SVS22330017SA, SVS22330018SA, SVS22330019SA, SVS22330020SA, SVS22330021SA, SVS22330022SA, SVS22330023SA, SVS22330024SA, SVS22330025SA, SVS22330026SA, SVS22330028SA, SVS22330030SA, SVS22330031SA, SVS22330032SA, SVS22330033SA, SVS22330034SA, SVS22330035SA, SVS22440001SA, SVS22440002SA, SVS22440005SA, SVS22440008SA, SVS22440009SA, SVS22440010SA, SVS22440011SA, SVS22440012SA, SVS22440013SA, SVS22440016SA, SVS22490002SA, SVS22490003SA, SVS22490004SA, SVS22520006SA, SVS22520008SA, SVS22520009SA, SVS22520010SA, SVS22520011SA, SVS22520012SA, SVS22520013SA, SVS22520015SA, SVS23040001SA, SVS23040002SA, SVS23040003SA, SVS23040004SA, SVS23040005SA, SVS23040007SA, SVS23040008SA, SVS23040011SA, SVS23040012SA, SVS23040013SA, SVS23040014SA, SVS23040015SA, SVS23040016SA, SVS23040017SA, SVS23040018SA, SVS23040019SA, SVS23070002SA, SVS23070004SA, SVS23070005SA, SVS23070008SA, SVS23070009SA, SVS23070011SA, SVS23070012SA, SVS23100001SA, SVS23100002SA, SVS23100004SA, and SVS23100006SA. Serial numbers for FRUs: SVS22330015SA, SVS22440006SA, SVS22440007SA, SVS22440015SA, SVS22490001SA, SVS22520005SA, and SVS22520014SA.

Product
Prucka 3 Amplifiers, Model P1801PA, and Field Replaceable Units (FRUs), Model 5875569, used with CardioLab AltiX and ComboLab AltiX.
Recalling firm
GE Medical Systems Information Technologies Inc
Quantity
81 amplifiers
Distribution
Distribution was made to CA, CO, FL, GA, IL, IN, KY, MA, NC, NJ, OH, OR, PA, SC, TX, VA, and WI. There was government distribution but no military distribution. Foreign distribution was made to Canada, France, Germany, Greece, Ireland, Japan, Republic of Korea, and Romania.
Recall date
September 27, 2023
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls