FDAClass IIJuly 8, 2026

Medline Convenience kits containing recalled Swan-Ganz

Medline Convenience kits containing recalled Swan-Ganz Catheters RIGHT HEART CATH PACK CTX DYNJ68530A

Why was Medline Convenience kits containing recalled Swan-Ganz recalled?

Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of leaks/breakage involving failures of the (blue) proximal injectate lumen hub on these catheters. This has led to instances of leakage, lumen damage, and breakage, and may lead to infection, medication loss, and/or blood loss.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

UDI-DI 10195327033620 Lots 21KBN891 21KBO410 22ABA676 22ABY937 22BBN909 22BBT447 23ABO303

Product
Medline Convenience kits containing recalled Swan-Ganz Catheters RIGHT HEART CATH PACK CTX DYNJ68530A
Recalling firm
Medline Industries, LP
Quantity
9,272 kits
Distribution
US Nationwide distribution in the states of AZ, CA, FL, IL, MO, NJ, OH, OK, TN, TX, WI.
Recall date
July 8, 2026
Classification
Class II
Status
Ongoing
Agency
FDA (device)