FDAClass IIJuly 8, 2026

Medline Convenience kits containing recalled Swan-Ganz

Medline Convenience kits containing recalled Swan-Ganz Catheters RIGHT HEART CATH PACK CTX DYNJ68530B

Why was Medline Convenience kits containing recalled Swan-Ganz recalled?

Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of leaks/breakage involving failures of the (blue) proximal injectate lumen hub on these catheters. This has led to instances of leakage, lumen damage, and breakage, and may lead to infection, medication loss, and/or blood loss.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

UDI-DI 10195327345044 Lots 23DBI807 23FBG708 23GBK642 23KBO477 24BBG198 24FBN889 24GBJ879 24HBR045 24LBT492 25ABL957 25BBN572 25GBT734 25HBQ817 25LBM617 26BBK766 26DBE720

Product
Medline Convenience kits containing recalled Swan-Ganz Catheters RIGHT HEART CATH PACK CTX DYNJ68530B
Recalling firm
Medline Industries, LP
Quantity
9,272 kits
Distribution
US Nationwide distribution in the states of AZ, CA, FL, IL, MO, NJ, OH, OK, TN, TX, WI.
Recall date
July 8, 2026
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls