MicroScan Pos Combo Panel Type 43 recall
MicroScan Pos Combo Panel Type 43, Catalog number B1017-217for use in determining antimicrobial agent susceptibility and/or identification to the…
Why was MicroScan Pos Combo Panel Type 43 recalled?
The drug dilution sequence for Moxifloxacin printed on the panel is incorrect. As a result, incorrect MIC results for Mxf could be reported.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Lot 2018-11-03
- Product
- MicroScan Pos Combo Panel Type 43, Catalog number B1017-217for use in determining antimicrobial agent susceptibility and/or identification to the…
- Recalling firm
- Beckman Coulter Inc.
- Quantity
- 686 boxes
- Distribution
- NJ, TX, IA, PA, FL, NE, OH, NY, CA, MO, OK, LA
- Recall date
- August 8, 2018
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)