Flow sensors that may be installed in the following recall
Flow sensors that may be installed in the following GEHC anesthesia machines: (1) Aestiva MRI, GTIN 0080682102339; (2) Aestiva 7900; (3) Aestiva 7100; (4) Aespire System Carestation, S/5 Protiva Config., Aespire 100 Compact Wall Mount (Aespire 7100); (5) S/5 Aespire 7900; (6) Aespire View; (7) Avance; (8) Avance CS2, GTIN 00840682102292; (9) Aisys; (10) Aisys CS2, GTIN 00840682102322; (11) Amingo; (12) 9100 NXT Config, Aelite NXT; (13) Protiva 7100;
Why was Flow sensors that may be installed in the following recalled?
The flow sensors built prior to June 2021 could have damaged tubes with small punctures or cuts that could cause leaks resulting in incorrect anesthesia machine tidal volumes potentially leading to over-delivery of volume to the patient.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
The flow sensors built prior to June 2021 that are installed in the devices/anesthesia machines.
- Product
- Flow sensors that may be installed in the following GEHC anesthesia machines: (1) Aestiva MRI, GTIN 0080682102339; (2) Aestiva 7900; (3) Aestiva 7100; (4) Aespire System Carestation, S/5 Protiva Config., Aespire 100 Compact Wall Mount (Aespire 7100); (5) S/5 Aespire 7900; (6) Aespire View; (7) Avance; (8) Avance CS2, GTIN 00840682102292; (9) Aisys; (10) Aisys CS2, GTIN 00840682102322; (11) Amingo; (12) 9100 NXT Config, Aelite NXT; (13) Protiva 7100;
- Recalling firm
- GE Healthcare, LLC
- Quantity
- Estimated 310,072 flow sensors
- Distribution
- Distribution was nationwide, including PR. Foreign distribution was made to Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belgium, Bermuda, Bolivia, Bosnia and Herzegovina, Brazil, Brunei, Bulgaria, Cambodia, Canada, Chile, China, Colombia, Costa Rica, Cote D'lvoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Fiji, Finland, France, Georgia, Germany, Ghana, Greece, Guatemala, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jersey, Jordan, Kazakhstan, Kenya, Korea, Kosovo, Kuwait, Laos, Latvia, Lebanon, Libya, Lithuania, Macedonia, Madagascar, Malaysia, Maldives, Mali, Malta, Mauritius, Mexico, Moldova, Montenegro, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Nicaragua, Niger, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Rwanda, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Trinidad and Tobago, Tunisia, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom Uruguay, Venezuela, Vietnam, and Yemen.
- Recall date
- October 6, 2021
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)