FDAClass IIAugust 24, 2016

Hospira Receptal Canisters: 1l(1000ml) recall

Hospira RECEPTAL CANISTERS: 1L(1000mL), 1.5L(1500mL), 2L(2000mL) Canister, ATS, 2000, with Valve 1) List Numbers 43449-01 and 43449-11; LIST NO. 43449…

Why was Hospira Receptal Canisters: 1l(1000ml) recalled?

Hospira has become aware of customers using the incorrect size Receptal liners with their Receptal canisters, e.g., 1.5 L liner, 2L liner or 3L liner with a 1L canister. It is possible that if an incorrectly sized liner is used with a different size canister the lid may not fit and the vacuum/suction will not be sufficient and/or consistent.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

43423-04-01 43423-04-11 43423-04-21 43445-04-01 43445-04-11 43449-04-01 43449-04-11 43496-04-01

Product
Hospira RECEPTAL CANISTERS: 1L(1000mL), 1.5L(1500mL), 2L(2000mL) Canister, ATS, 2000, with Valve 1) List Numbers 43449-01 and 43449-11; LIST NO. 43449…
Recalling firm
Hospira Inc.
Quantity
67,410,861 liners and 67,410,861 lids/canisters in total
Distribution
*** US: Nationwide (AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY), District of Columbia, Puerto Rico; *** FOREIGN: Australia, Canada, Costa Rica, Singapore, South Africa
Recall date
August 24, 2016
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls