FDAClass IIJune 28, 2017

Proteus 235 recall

Proteus 235

Why was Proteus 235 recalled?

It has been observed during testing that there could be clinically non-negligible amount of dark current (a well-known physics effect observed in cyclotron) in Pencil Beam Scanning (PBS) treatment mode when system is not irradiating between layers.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Serial Numbers: PAT.000,PAT.003, PAT.006, PAT.107, PAT.108, PAT.109, PAT.111, PAT.112, PAT.113, PAT.114, SAT.115, SAT.116, SAT.117, SAT.119, SAT.120, SAT.122, SAT.123, SAT.126

Product
Proteus 235
Recalling firm
Ion Beam Applications S.A.
Quantity
18 systems ( 10 in the US)
Distribution
Worldwide Distribution - US Distribution to the states of : MA, FL, VA, PA, IL, NJ, WA, TN, LA and TX., and to the countries of: Germany, South Korea, France, Czech Republic, Italy, Poland and Sweden.
Recall date
June 28, 2017
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls