FDAClass IAugust 28, 2024

ClotTriever XL Catheter recall

ClotTriever XL Catheter, 30 mm, REF: 41-102, used with Instructions for Use ClotTriever XL Catheter ClotTriever Thrombectomy System, Reference…

Why was ClotTriever XL Catheter recalled?

Thrombectomy system catheter instructions for use updated to reduce adverse events due to the following circumstances: 1) Fibrous, firmly adherent, calcified material, or suspected tumor thrombus overloading catheter bag 2) Catheter pulled through internal jugular vein access site or not pulled away from heart; may lead to device entrapment, pulmonary embolism, venous cutdown, abdominal surgery.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

UDI-DI:00850291007277. All lots. IU-01011 REV. C.

Product
ClotTriever XL Catheter, 30 mm, REF: 41-102, used with Instructions for Use ClotTriever XL Catheter ClotTriever Thrombectomy System, Reference…
Recalling firm
Inari Medical - Oak Canyon
Quantity
2488
Distribution
Worldwide - US Nationwide distribution including in the states of PA, FL, GA, CA, IL, SC, NY, MN, UT, TN, WI, MI, KS, SD, AZ, AR, IN, KY, TX, AL, MS, DE, MA, OH, MO, WV, VA, NC, NV, CO, NE, OR, NM, NH, NJ, CT, ID, IA, WA, MD, OK, RI, DC, PR, LA, ND, VT and the countries of New Zealand, Chile, Australia.
Recall date
August 28, 2024
Classification
Class I
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls