FDAClass IIOctober 6, 2021

AtomLab 500Plus Dose CalibratorSoftware 2.0.00 through recall: Software Issue

AtomLab 500Plus Dose CalibratorSoftware 2.0.00 through Revision 2.0.08, inclusive Model: 086-332

Why was AtomLab 500Plus Dose CalibratorSoftware 2.0.00 through recalled?

When deleting a previously entered custom isotope, the software deletes the isotope, but not the associated dial setting. Possible other stored custom isotopes may then have incorrect dial settings and may lead to an incorrect dose measurement and/or misadministration

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Serial Number Range:17080651-21061047 UDI: 00718175003329

Product
AtomLab 500Plus Dose CalibratorSoftware 2.0.00 through Revision 2.0.08, inclusive Model: 086-332
Recalling firm
Biodex Medical Systems, Inc.
Quantity
429 units
Distribution
Nationwide Foreign: Austria Autralia Brazil CANADA Chile Colombia Croatia Cyprus Czech Republic Denmark Dominican Republic Ecuador Egypt France Hong Kong HUNGARY Iceland India Iraq Ireland ISRAEL Isreal Italy Japan JORDAN Kuwait Lebanon Libya Malaysia MALTA Mexico Morocco Netherlands New Zealand Oman Pakistan Palestine Panama Peru Philippines Poland Portugal PR Puerto Rico Qatar Romania Russia Saudi Arabia Singapore South Africa South Korea Spain Sweden Taiwan UAE United Arab Emirates United Kingdom Venezuela
Recall date
October 6, 2021
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls