FDAClass IIAugust 8, 2018

Liquid Cardiac Control - Level 3 recall

Liquid Cardiac Control - Level 3, cat. no. CQ5053 Product Usage: This product is intended for in vitro diagnostic use, in the quality control of…

Why was Liquid Cardiac Control - Level 3 recalled?

The device is not suitable for the control of the Troponin T assay due to unacceptable variation between vials. The deviation in precision observed is up to 28%. IQC that is reported as out of range could lead to a delay in reporting Troponin T results. If the failure occurs the operator may observe the quality control result for Troponin T falling outside range and/or increased imprecision for Quality Control monitoring.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Lot 4245CK.

Product
Liquid Cardiac Control - Level 3, cat. no. CQ5053 Product Usage: This product is intended for in vitro diagnostic use, in the quality control of…
Recalling firm
Randox Laboratories Ltd.
Quantity
1 kit (US), 602 kits (OUS)
Distribution
Worldwide Distribution - US in the state of WV. Foreign distribution to Australia, Chile, China, Germany, Hungary, Ireland, Italy, Moldova, Netherlands, Peru, Poland, Saudi Arabia, Slovakia, Spain, Sultanate of Oman, UAE, UK, and Vietnam
Recall date
August 8, 2018
Classification
Class II
Status
Terminated
Agency
FDA (device)