NanoDots recall
NanoDots, Model Numbers: a) BC30023, description: QC Set For Microstar Dots; b) BC30083, description: nanoDot QC Set - 80kVp - Sale; c) BC30084…
Why was NanoDots recalled?
LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
UDI/DI 0860003399903, all batch numbers
- Product
- NanoDots, Model Numbers: a) BC30023, description: QC Set For Microstar Dots; b) BC30083, description: nanoDot QC Set - 80kVp - Sale; c) BC30084…
- Recalling firm
- Landauer
- Quantity
- 18333 units
- Distribution
- Worldwide
- Recall date
- September 20, 2023
- Classification
- Class II
- Status
- Completed
- Agency
- FDA (device)