FDAClass IISeptember 20, 2023

Constancy recall

Constancy (80 kVp) nanoDot D2DNN, Model Numbers: a) 18150-000 (EMEA number VKITCON002); b) 18150-KIT; radiation monitoring dosimeter used with the…

Why was Constancy recalled?

LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

UDI/DI 0860003399903, all batch numbers

Product
Constancy (80 kVp) nanoDot D2DNN, Model Numbers: a) 18150-000 (EMEA number VKITCON002); b) 18150-KIT; radiation monitoring dosimeter used with the…
Recalling firm
Landauer
Quantity
512 units
Distribution
Worldwide
Recall date
September 20, 2023
Classification
Class II
Status
Completed
Agency
FDA (device)
Open FDA enforcement record Latest recalls