Roche cobas IT 1000 Product Usage: Intended Use: An recall
Roche cobas IT 1000 Product Usage: Intended Use: An electronic device intended to store, retrieve, and process laboratory data.
Why was Roche cobas IT 1000 Product Usage: Intended Use: An recalled?
Potential that test results would be assigned to the wrong Patient ID if cobas IT 1000 receives an ADT ADMIT, TRANSFER or MERGE event message that contains a genuine Patient ID (e.g. 999999), which is the same as what has been configured as the Generic Patient ID (999999 by default).
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Part Numbers: 04420306001 Serial #'s: 2706, 2724, 2743, 2744, 2745, 2749, 2750, 2751, 2752, 2754, 2755, 2764, 2770, 2771, 2772, 2773, 2774, 2775, 2776, 2780, 2781, 2782, 2783, 2784, 2785, 2789, 2790, 2791, 2792, 2793, 2794, 2795, 2796, 2797, 2798, 2799, 2808, 2814, 2815, 2816, 2818, 2819, 2823, 2824, 2825, 2826, 2828, 2829, 2830, 2831, 2847, 2848, 2849, 2857, 2858, 2859, 2860, 2861, 2862, 2863, 2864, 2865, 2904, 2905, 2906, 2908, 2909, 2914, 2915, 2916, 2917, 2920, 2921, 2928, 2931, 2932, 2933, 2934, 2935, 2936, 2937, 2938, 2939, 2940, 2941, 2943, 2983, 2984, 2985, 2986, 2987, 2990, 2991, 2992, 2993, 2994, 2995, 2996, 2997, 2998, 3006, 3007, 3008, 3009, 3016, 3020, 3021, 3023, 3024, 3025, 3026, 3027, 3048, 3049, 3050, 3051, 3053, 3057, 3058, 3059, 3060, 3061, 3065, 3066, 3067, 3068, 3069, 3070, 3071, 3072, 3073, 3074, 3075, 3076, 3077, 3079, 3082, 3083, 3084, 3085, 3086, 3087, 3089, 3090, 3104, 3105, 3106, and 3107.
- Product
- Roche cobas IT 1000 Product Usage: Intended Use: An electronic device intended to store, retrieve, and process laboratory data.
- Recalling firm
- Roche Diagnostics Operations, Inc.
- Quantity
- 148 Units
- Distribution
- US Nationwide Distribution in the states of: AK, AL, AR, AZ, CA, CT, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MD, ME, MI, MO, NC, NE, NH, NJ, NY, OH, OR, PA, PR, SC, TN, TX, VA, VT, WA, WI, and WV.
- Recall date
- September 10, 2014
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)