FDAClass IISeptember 24, 2025

Chemistry Products LAC Slides recall

Chemistry Products LAC Slides. Catalog Number: 8433880. 1 Sales Unit = contains 5 cartridges of 60 slides, for a total of 300 tests.

Why was Chemistry Products LAC Slides recalled?

VITROS Chemistry Products LAC Slides on VITROS XT 3400 and VITROS XT 7600 Systems are experiencing an increased occurrence of calibration failure related to lots from coating 0130 and above with Condition Code TH4-63J during calibration, causing the delay of lac test results and potentially impact patient management. The issue was identified by QuidelOrtho through an increase in customer complaints compared to the historical baseline with average failure rate of 1% for coating 0130 in 2025 while average of failure rate for other coating was 0.06% in 2025 and average of failure rate for all coatings is 0.03% for 2024.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Catalog number 8433880; UDI# 110758750004911; Coating 0130 and above, this includes Lot numbers 3533-0130-2317, 3533-0130-2717, 3533-0130-2718, 3533-0130-2719, 3533-0130-4546, 3533-0130-4605, 3533-0131-4763, 3533-0131-6114, 3533-0131-6115, 3533-0131-6116, 3533-0131-6128; Expiration date range 01-Sep-2026 to 01-Jan-2027

Product
Chemistry Products LAC Slides. Catalog Number: 8433880. 1 Sales Unit = contains 5 cartridges of 60 slides, for a total of 300 tests.
Recalling firm
Ortho-Clinical Diagnostics, Inc.
Quantity
2,908 units
Distribution
US Nationwide distribution.
Recall date
September 24, 2025
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls