Cordis EMPIRA recall
Cordis EMPIRA RX PTCA Dilatation Catheter
Why was Cordis EMPIRA recalled?
The balloon may exhibit resistance to the removal of the balloon cover. Weakening of the affected product in the region of the proximal balloon seal may occur when excessive force is applied to remove the balloon cover. This may lead to balloon burst, shaft separation or deflation difficulties, which could result in procedural delay, vessel trauma or surgery.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
70 Catalog numbers: 85R06150S, 85R06200S, 85R06225S, 85R06250S, 85R06275S, 85R06300S, 85R06325S, 85R06350S, 85R06375S, 85R06400S, 85R10150S, 85R10200S, 85R10225S,85R10250S, 85R10275S, 85R10300S, 85R10325S, 85R10350S, 85R10375S, 85R10400S, 85R12150S, 85R12200S, 85R12225S, 85R12250S, 85R12275S, 85R12300S, 85R12325S, 85R12350S, 85R12375S, 85R12400S, 85R15150S, 85R15200S, 85R15225S, 85R15250S, 85R15275S, 85R15300S, 85R15325S, 85R15350S, 85R15375S, 85R15400S, 85R20150S, 85R20200S, 85R20225S, 85R20250S, 85R20275S, 85R20300S, 85R20325S, 85R20350S, 85R20375S, 85R20400S, 85R25150S, 85R25200S, 85R25225S, 85R25250S, 85R25275S, 85R25300S, 85R25325S, 85R25350S, 85R25375S, 85R25400S, 85R30150S, 85R30200S, 85R30225S, 85R30250S, 85R30275S, 85R30300S 85R30325S, 85R30350S, 85R30375S, and 85R30400S.
- Product
- Cordis EMPIRA RX PTCA Dilatation Catheter
- Recalling firm
- Cordis Corporation
- Quantity
- 250,000 units.
- Distribution
- Worldwide Distribution - USA including Puerto Rico and the states of AR, AZ, CA, CO, CT, DC, FL, GA, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WV, WY and Hawaii., and the countries of Argentina, Brazil, Canada, Chile, Colombia, Mexico, Panama and Venezuela.
- Recall date
- September 3, 2014
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)