CURBELL MEDICAL CareSense Wireless Advanced LCD Bed and recall
CURBELL MEDICAL CareSense Wireless Advanced LCD Bed and Chair Monitor, MODEL # CSM-BC400; --- COMMON/USUAL NAME: Bed Patient Monitor; CLASSIFICATION…
Why was CURBELL MEDICAL CareSense Wireless Advanced LCD Bed and recalled?
The firm became aware of a potential problem that was initiated by a customer complaint. After consultation with the manufacturer, it was discovered that a resistor was incorrectly placed within the circuit board on the monitor. This change to the resistor was a planned change to address a product improvement (improve battery drain). However, this modification rendered the monitor's alarm via the nurse call communication port to not function.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Lot 0414
- Product
- CURBELL MEDICAL CareSense Wireless Advanced LCD Bed and Chair Monitor, MODEL # CSM-BC400; --- COMMON/USUAL NAME: Bed Patient Monitor; CLASSIFICATION…
- Recalling firm
- Curbell Medical, Inc.
- Quantity
- Domestic: 112 units; Foreign: 211 units
- Distribution
- Nationwide, Canada, Ireland, Italy
- Recall date
- September 3, 2014
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)