Trauma Tubes and Plates: Tube & SCP PLT 90dx150lg Tube recall
Trauma Tubes and Plates: TUBE & SCP PLT 90DX150LG TUBE & SCP PLT 90DX205LG TUBE & SCP PLT 90DX255LG TUBE & SCP PLT 95DX150LG TUBE & SCP PLT 95DX205LG…
Why was Trauma Tubes and Plates: Tube & SCP PLT 90dx150lg Tube recalled?
Zimmer is recalling sterile-packaged implants and instruments packaged in six different package configurations due to packaging design verification test failures. Specifically, multiple test samples from each of the six configurations failed simulated distribution and shipping testing when the devices either breached inner or outer package trays or seal integrity was lost.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Item Numbers: 00118109006, 00118109008, 00118109010, 00118109506, 00118109508, 00118109510, 00118113008, 00118113010, 00118113012, 00118113014, 00118113508, 00118113510, 00118113512, 00118113514, 00118114008, 00118114010, 00118114012, 00118114014, 00118114508, 00118114510, 00118114512, 00118114514, 00118115006, 00118115008, 00118115010, 00118115012, 00118115014, 00119309006, 00119309008, 00119309010, 00119309012, 00119309014, 00119309506, 00119309508, 00119309510, 00119309512, 00119309514, 00119809008, 00119809011, 00119809508, 00119809511, 00119809514, 00119313008, 00119313010, 00119313012, 00119313014, 00119313508, 00119313510, 00119313512, 00119313514, 00119314008, 00119314010, 00119314012, 00119314014, 00119314508, 00119314510, 00119314512, 00119314514, 00119315008, 00119315010, 00119315012, 00119315014, 00119413008, 00119413508, 00119414008, 00119414508, 00119415008, 00119809008, 00119809011, 00119809508, 00119809511, 00119809514 EXPIRATION DATE PRIOR TO 2024-07
- Product
- Trauma Tubes and Plates: TUBE & SCP PLT 90DX150LG TUBE & SCP PLT 90DX205LG TUBE & SCP PLT 90DX255LG TUBE & SCP PLT 95DX150LG TUBE & SCP PLT 95DX205LG…
- Recalling firm
- Zimmer, Inc.
- Quantity
- 972,125 units
- Distribution
- Worldwide Distribution.
- Recall date
- September 3, 2014
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)