FDAClass IISeptember 3, 2014

Trauma Large Screws: Cann Bone Screw 7.0mmx100mmlg Cann recall

Trauma Large Screws: CANN BONE SCREW 7.0MMX100MMLG CANN BONE SCREW 7.0X105LG CANN BONE SCREW 7.0X105LG CANN BONE SCREW 7.0MMX105MMLG CANN BONE SCREW…

Why was Trauma Large Screws: Cann Bone Screw 7.0mmx100mmlg Cann recalled?

Zimmer is recalling sterile-packaged implants and instruments packaged in six different package configurations due to packaging design verification test failures. Specifically, multiple test samples from each of the six configurations failed simulated distribution and shipping testing when the devices either breached inner or outer package trays or seal integrity was lost.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Item Numbers 00114610099, 00114610500, 00114610532, 00114610599, 00114611000, 00114611032, 00114611099, 00114611500, 00114611532, 00114611599, 00114612000, 00114612032, 00114612099, 00114612500, 00114612599, 00114613000, 00114613099, 00225210555, 00225211055, 00225211555, 00225212055, 00225212555, 00225213055, 00225710565, 00225711065, 47234712005, 47234712400, 47234712440, 47234712500, 47234712600, 47234810535, 47234811035, 47235910055, 47248310560, 47248311060, 47248311560, 47248312060, 47248312560, 47248313060, 47248313560, 47248314060, 47248410560, 47248411060, 47248411560, 47248412060, 47248412560, 47248413060, 47248413560, 47248414060, 47348310560, 47348311060, 47348311560, 47348312060, 47348312560, 47348313060, 47348313560, 47348314060, 47348410560, 47348411060, 47348411560, 47348412060, 47348412560, 47348413060, 47348413560, 47348414060, 47483510500, 47483510501, 47483511000, 47483511001, 47483511500, 47483511501, 47483512000, 47483512001, 47483512500, 47483513000, 47483513500, 47483514000, 47483514500, 47483515000, 47484510501, 47484511001, 47484512001, 47486510500, 47486510501, 47486510502, 47486511000, 47486511001, 47486511002, 47486511502, 47486512002 EXPIRATION DATE PRIOR TO 2024-07

Product
Trauma Large Screws: CANN BONE SCREW 7.0MMX100MMLG CANN BONE SCREW 7.0X105LG CANN BONE SCREW 7.0X105LG CANN BONE SCREW 7.0MMX105MMLG CANN BONE SCREW…
Recalling firm
Zimmer, Inc.
Quantity
972,125 units
Distribution
Worldwide Distribution.
Recall date
September 3, 2014
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls