FDAClass IISeptember 3, 2014

Endo & Unipolar Femoral Heads: Endo Femoral Head 41mm recall

Endo & Unipolar Femoral Heads: ENDO FEMORAL HEAD 41MM DIA ENDO FEMORAL HEAD 42MM DIA ENDO FEMORAL HEAD 43MM DIA ENDO FEMORAL HEAD 44MM DIA ENDO…

Why was Endo & Unipolar Femoral Heads: Endo Femoral Head 41mm recalled?

Zimmer is recalling sterile-packaged implants and instruments packaged in six different package configurations due to packaging design verification test failures. Specifically, multiple test samples from each of the six configurations failed simulated distribution and shipping testing when the devices either breached inner or outer package trays or seal integrity was lost.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Item Numbers 00781804100, 00781804200, 00781804300, 00781804400, 00781804500, 00781804600, 00781804700, 00781804800, 00781804900, 00781805000, 00781805100, 00781805200, 00781805300, 00781805400, 00781805500, 00781805700, 00781806000, 00781806300, 00902604101, 00902604201, 00902604301, 00902604401, 00902604501, 00902604601, 00902604701, 00902604801, 00902604901, 00902605001, 00902605101, 00902605201, 00902605301, 00902605401, 00902605501, 00902605701, 00902606001, 00902606301, 451000041, 451000042, 451000043, 451000044, 451000045, 451000046, 451000047, 451000048, 451000049, 451000050, 451000051, 451000052, 451000053, 451000054, 451000055, 451000056, 451000057, 451000058, 451000059, 451000060 EXPIRATION DATE PRIOR TO 2024-01

Product
Endo & Unipolar Femoral Heads: ENDO FEMORAL HEAD 41MM DIA ENDO FEMORAL HEAD 42MM DIA ENDO FEMORAL HEAD 43MM DIA ENDO FEMORAL HEAD 44MM DIA ENDO…
Recalling firm
Zimmer, Inc.
Quantity
972,125 units
Distribution
Worldwide Distribution.
Recall date
September 3, 2014
Classification
Class II
Status
Terminated
Agency
FDA (device)