FDAClass IISeptember 3, 2014

Knee Instrument Fixation Screws: 48MM HEADED SCREW 48MM recall

Knee Instrument Fixation Screws: 48MM HEADED SCREW 48MM HEADLESS SCREW MIS HEADED SCREW, 48MM MIS HEADLESS SCREW, 48MM

Why was Knee Instrument Fixation Screws: 48MM HEADED SCREW 48MM recalled?

Zimmer is recalling sterile-packaged implants and instruments packaged in six different package configurations due to packaging design verification test failures. Specifically, multiple test samples from each of the six configurations failed simulated distribution and shipping testing when the devices either breached inner or outer package trays or seal integrity was lost.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Item Numbers 00579104100, 00579104200, 00598304048, 00598304148 Exp Date Prior to 2023-11

Product
Knee Instrument Fixation Screws: 48MM HEADED SCREW 48MM HEADLESS SCREW MIS HEADED SCREW, 48MM MIS HEADLESS SCREW, 48MM
Recalling firm
Zimmer, Inc.
Quantity
972,125 units
Distribution
Worldwide Distribution.
Recall date
September 3, 2014
Classification
Class II
Status
Terminated
Agency
FDA (device)