FDAClass IIJuly 1, 2026

Off-Axis Comprehensive Shoulder System Small recall

Off-Axis Comprehensive Shoulder System Small, Augment, Off-Axis, Reamer Guide, Model/Catalog Number: 110040240

Why was Off-Axis Comprehensive Shoulder System Small recalled?

Five complaints have been received identifying issues where the plastic guides were catching during reaming, leading to breakage.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Lot Code: Model No. 110040240 ; UDI-DI (01)00889024681965(17)350601(10)099457 ; Lot Number 099457 Model No. 110040240 ; UDI-DI (01)00889024681965(17)350712(10)099458 ; Lot Number 099458

Product
Off-Axis Comprehensive Shoulder System Small, Augment, Off-Axis, Reamer Guide, Model/Catalog Number: 110040240
Recalling firm
Zimmer, Inc.
Quantity
474 units
Distribution
US Nationwide distribution.
Recall date
July 1, 2026
Classification
Class II
Status
Ongoing
Agency
FDA (device)