Artis zee systems with Laird Cooling Unit for SSFD recall
Artis zee systems with Laird Cooling Unit for SSFD: Material # 7555118. Artis is an angiography system developed for single and biplane diagnostic…
Why was Artis zee systems with Laird Cooling Unit for SSFD recalled?
Due to a defective sealing, coolant may inflow into the electrical parts of the system cabinet, which could cause a short circuit and cause the system to shut down suddenly during an ongoing procedure
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Laird Cooling Unit for SSFD: Material # 7555118.
- Product
- Artis zee systems with Laird Cooling Unit for SSFD: Material # 7555118. Artis is an angiography system developed for single and biplane diagnostic…
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Quantity
- 3 US
- Distribution
- HI , IL, MI, NC, NY, VA, WI
- Recall date
- August 1, 2018
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)