FDAClass IIJuly 15, 2020

uCT 760 System recall

uCT 760 System, Computed Tomography X-ray System

Why was uCT 760 System recalled?

Two issues were identified with the computed tomography x-ray system including an intermittent issue which may affect image data leading to artifacts; and intermittent scout scanning interruption due to occasional angle signal drift. This may result in the need for rescanning of the patient with the patient receiving an additional dose of radiation and possible use of additional contrast media.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Model uCT 760; Serial Numbers: 600229, 600230, 600232, 600271 with software version R001.11.0.729741-Re-1100, and serial number 600282 with software version R001.11.0.785930-Re-1150.

Product
uCT 760 System, Computed Tomography X-ray System
Recalling firm
Shanghai United Imaging Healthcare Co., Ltd.
Quantity
5 units
Distribution
Distributed to two customers. Both customers located in Houston, Texas.
Recall date
July 15, 2020
Classification
Class II
Status
Terminated
Agency
FDA (device)