FDAClass IIJune 24, 2026

Brand Name: Inspire Medical Systems recall: Impurity Out of Spec

Brand Name: Inspire Medical Systems, Inc. Product Name: Inspire stimulation Lead Model/Catalog Number: 4063 Product Description: The Inspire¿…

Why was Brand Name: Inspire Medical Systems recalled?

Products are labeled with incorrect Use By Date, extending beyond the validated three-year shelf life from date of manufacturer. Material degradation of a product past its Use By Date may result in unplanned or premature revision surgery.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Lot Code: Model 4063 (USA only lead); UDI-DI [4063] 10855728005793; Inspire part number [4063] 900-013-001; serial numbers D53393, D61104 Model 4063 (Global lead); UDI-DI [4063] 10810098650295, Inspire part number [4063] 900-013-003; serial numbers D89539, D89540, D89543, D89559, D89563, D89564, D89577, D8959

Product
Brand Name: Inspire Medical Systems, Inc. Product Name: Inspire stimulation Lead Model/Catalog Number: 4063 Product Description: The Inspire¿…
Recalling firm
Inspire Medical Systems Inc.
Quantity
10 impacted units
Distribution
Worldwide distribution - US Nationwide and the country of Hong Kong.
Recall date
June 24, 2026
Classification
Class II
Status
Ongoing
Agency
FDA (device)