FDAClass IIAugust 14, 2024

TruDi NAV Suction recall

TruDi NAV Suction, 0 Degrees; Model/Catalog Number: TDNS000Z. TruDi" NAV Suction Instruments are intended for use with the TruDi" Navigation System…

Why was TruDi NAV Suction recalled?

Specific lots of TruDi NAV Suction Instruments 0 Degrees were incorrectly calibrated which when used with the TruDi Navigation System, may result in a discrepancy between the actual tip location and the visualized location displayed on the system.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

UDI/DI: 10705031245877; Lot Codes: 2310160, 2309110, 2309150, 2308140, 2308160, 2308180.

Product
TruDi NAV Suction, 0 Degrees; Model/Catalog Number: TDNS000Z. TruDi" NAV Suction Instruments are intended for use with the TruDi" Navigation System…
Recalling firm
Integra LifeSciences Corp.
Quantity
265 units
Distribution
The devices are only sold in the US. The following states have received the product: AL, AZ CA, CO, CT, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NJ, NM, NY, OH, PA, SC, SD, TN, TX, VA, VT, WI.
Recall date
August 14, 2024
Classification
Class II
Status
Ongoing
Agency
FDA (device)