Visualase Cooled Laser Applicator System recall
Visualase Cooled Laser Applicator System (VCLAS) labeled as: a. Part Number 9735559, Visualase¿ Cooled Laser Applicator System 3mm Tip; b. Part Number…
Why was Visualase Cooled Laser Applicator System recalled?
Medtronic has become aware that certain lots of the VCLAS 3MM, 10MM, and 15MM laser diffusing fibers were not tested at intended laser power levels, due to an issue related to calibration of the test equipment, which can lead to the potential for low laser energy output from fiber or fiber burnout at lower wattage.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
a. Part Number 9735559, UDI (00643169611986, 00643169933361), Lot Codes (0212782059, 0212894256, 0213251294, 0213331774, 0213383266, 0212783784, 0212926008, 0213259125, 0213335148, 0213383267, 0212823883, 0212956163, 0213299882, 0213375479, 0213387215, 0212823919, 0212978141, 0213299885, 0213378982, 0213537710, 0212854585, 0212996201, 0213327285, 0213383244, 0213966003, 0212882619, 0213519510); b. Part Number 9735560, UDI (00643169611993, 00643169933385), Lot Codes (0212086315, 0213158249, 0213299883, 0213427083, 0213728326, 0212698642, 0213158631, 0213304570, 0213455321, 0213735817, 0212715926, 0213158719, 0213304571, 0213458938, 0213737565, 0212745633, 0213165510, 0213304573, 0213507022, 0213760863, 0212753561, 0213165511, 0213308717, 0213507023, 0213760864, 0212756095, 0213205023, 0213309087, 0213512428, 0213760866, 0212760821, 0213205024, 0213334598, 0213512430, 0213769912, 0212827856, 0213208868, 0213368255, 0213513054, 0213917334, 0212838837, 0213212692, 0213368499, 0213513055, 0213922188, 0212843810, 0213223891, 0213368573, 0213513056, 0213924407, 0212846727, 0213223892, 0213373657, 0213540031, 0213924418, 0212856596, 0213242839, 0213374559, 0213576515, 0213928892, 0212873863, 0213245723, 0213375535, 0213576516, 0213945813, 0212873873, 0213258625, 0213375537, 0213576517, 0213945814, 0212879480, 0213259039, 0213375538, 0213594271, 0213966370, 0212913269, 0213259673, 0213375991, 0213594272, 0213972422, 0212922387, 0213259674, 0213376135, 0213594273, 0213973527, 0212935278, 0213267742, 0213376136, 0213596471, 0214032331, 0213000097, 0213289058, 0213376138, 0213699385, 0214043408, 0213001255, 0213289059, 0213376139, 0213699386, 0214045295, 0213009483, 0213292765, 0213376655, 0213699390, 0214049428, 0213041517, 0213292893, 0213376656, 0213710007, 0214062101, 0213044327, 0213296890, 0213376780, 0213710711, 0214066559, 0213056030, 0213296891, 0213377536, 0213710712, 0214068011, 0213107317, 0213298978, 0213378061, 0213720928, 0214068013, 0213116115, 0213299342, 0213387201, 0213720929, 0214068921, 0213134866, 0213299525, 0213388160, 0213728324, 0214068922, 0213136534, 0213518754, 0213138278); c. Part Number 9735561, UDI (00643169612006, 00763000015695, 00643169933378), Lot Codes (0212701679, 0212912501, 0212991152, 0213647807, 0213669275, 0212707035, 0212935757, 0213555835, 0213662428, 0213669402, 0212745660, 0212939348, 0213555836, 0213663333, 0213669408, 0212796559, 0212963182, 0213645747, 0213663500, 0213670155, 0212846726, 0212965542, 0213645760, 0213667299, 0213670722, 0212901580, 0212904407)
- Product
- Visualase Cooled Laser Applicator System (VCLAS) labeled as: a. Part Number 9735559, Visualase¿ Cooled Laser Applicator System 3mm Tip; b. Part Number…
- Recalling firm
- Medtronic Navigation, Inc.
- Quantity
- 1,283 devices total
- Distribution
- Worldwide Distribution - US Nationwide in the states of AL, AR, AZ, CA, CO, DC, FL, GA, IL, IN, KY, LA, MA, MD, MI, MN, MO, NC, NJ, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI. International distribution to Israel.
- Recall date
- July 25, 2018
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)