FDAClass IIJune 24, 2026

MEDLINE Medical Procedure Kits labeled as: 1) CAROTID recall

MEDLINE Medical Procedure Kits labeled as: 1) CAROTID NEURO PACK, Medline Kit Number/SKU DYNJ44926G; 2) VP SHUNT CDS, Medline Kit Number/SKU…

Why was MEDLINE Medical Procedure Kits labeled as: 1) CAROTID recalled?

During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Medline Kit Number/SKU DYNJ44926G: UDI/DI 10193489766554 (each), 40193489766555 (case), Lot Number 24EBH848; Medline Kit Number/SKU DYNJ44926G: UDI/DI 10193489766554 (each), 40193489766555 (case), Lot Number 24DBK995; Medline Kit Number/SKU DYNJ44926G: UDI/DI 10193489766554 (each), 40193489766555 (case), Lot Number 23JBP023; Medline Kit Number/SKU DYNJ44926G: UDI/DI 10193489766554 (each), 40193489766555 (case), Lot Number 23HBR067; Medline Kit Number/SKU DYNJ44926G: UDI/DI 10193489766554 (each), 40193489766555 (case), Lot Number 23FBM330; Medline Kit Number/SKU DYNJ44926G: UDI/DI 10193489766554 (each), 40193489766555 (case), Lot Number 23EBK061; Medline Kit Number/SKU DYNJ44926G: UDI/DI 10193489766554 (each), 40193489766555 (case), Lot Number 22JBG042; Medline Kit Number/SKU DYNJ44926G: UDI/DI 10193489766554 (each), 40193489766555 (case), Lot Number 22GBH864; Medline Kit Number/SKU DYNJ44926G: UDI/DI 10193489766554 (each), 40193489766555 (case), Lot Number 22DBT571; Medline Kit Number/SKU CDS983800J: UDI/DI 10195327684662 (each), 40195327684663 (case), Lot Number 25DMJ438; Medline Kit Number/SKU CDS983800J: UDI/DI 10195327684662 (each), 40195327684663 (case), Lot Number 25CMH178; Medline Kit Number/SKU CDS983800J: UDI/DI 10195327684662 (each), 40195327684663 (case), Lot Number 24IME720; Medline Kit Number/SKU CDS983800J: UDI/DI 10195327684662 (each), 40195327684663 (case), Lot Number 24HMG912; Medline Kit Number/SKU CDS983800K: UDI/DI 10198459518973 (each), 40198459518974 (case), Lot Number 26CMC503; Medline Kit Number/SKU CDS983800K: UDI/DI 10198459518973 (each), 40198459518974 (case), Lot Number 26AMB047; Medline Kit Number/SKU CDS983800K: UDI/DI 10198459518973 (each), 40198459518974 (case), Lot Number 25JME109.

Product
MEDLINE Medical Procedure Kits labeled as: 1) CAROTID NEURO PACK, Medline Kit Number/SKU DYNJ44926G; 2) VP SHUNT CDS, Medline Kit Number/SKU…
Recalling firm
Medline Industries, LP
Quantity
342 units
Distribution
Worldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI.
Recall date
June 24, 2026
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls