MEDLINE Medical Procedure Kits labeled as: DR FERNANDEZ recall
MEDLINE Medical Procedure Kits labeled as: DR FERNANDEZ, Medline kit number/SKU DYNJ68808
Why was MEDLINE Medical Procedure Kits labeled as: DR FERNANDEZ recalled?
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
UDI/DI 10193489843187 (each), 40193489843188 (case), Lot Numbers: 21LBT965, 21IBU081, 21GBN159, 21DBC976.
- Product
- MEDLINE Medical Procedure Kits labeled as: DR FERNANDEZ, Medline kit number/SKU DYNJ68808
- Recalling firm
- Medline Industries, LP
- Quantity
- 91 units
- Distribution
- Worldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI.
- Recall date
- June 24, 2026
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)