FDAClass IISeptember 9, 2015

RAPIDPoint 500 Blood Gas Analyzer recall

RAPIDPoint 500 Blood Gas Analyzer, Siemens Material Numbers (SMN):10492730, 10696855, 10696857, 10697306 This system tests blood gases, electrolytes…

Why was RAPIDPoint 500 Blood Gas Analyzer recalled?

When both ports (Serial and Ethernet) are configured to transmit data, the data stream from one port could potentially affect the data stream from the other port causing the data management system to include either: duplicate data, missing data, or data from a different patient.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Units shipped between the dates Oct 2011 - June 2015 and within the serial range of 30003-37040 not inclusive.

Product
RAPIDPoint 500 Blood Gas Analyzer, Siemens Material Numbers (SMN):10492730, 10696855, 10696857, 10697306 This system tests blood gases, electrolytes…
Recalling firm
Siemens Healthcare Diagnostics Inc
Quantity
5241 units
Distribution
Worldwide Distribution - US (nationwide) Distribution and to the countries of : Albania, Algeria, Armenia, Australia, Austria, Bahrain, Bangladesh, Belgium, Bosnia Herzeg, Brazil, Bulgaria,Canada, Chile, Colombia, Croatia, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Georgia, Germany,Greece, Hong Kong, Hungary, India, Indonesia, Ireland, Italy, Japan, Kenya, Kuwait, Latvia, Lebanon, Lithuania, Luxembourg, Malaysia, Mexico, Netherlands. New Zealand, Norway, P.R. China, Peru, Poland, Portugal, Qatar, Republic Korea, Romania, Russian Fed, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Tunisia, Turkey, U.A.E., United Kingdom and Vietnam.
Recall date
September 9, 2015
Classification
Class II
Status
Terminated
Agency
FDA (device)